Table 2. Mean blood pressure (BP) values obtained with two devices: reference auscultatory device always in correct position and test device (either auscultatory or oscillometric) in varying positions.

  Systolic BP
Diastolic BP
Position of test device cuff Reference Test Reference Test
Test device: auscultatory, appropriately sized cuff (Main Study; n=57)
 Correct 115.9±18.3 116.3±18.5 75.8±11.3 76.1±11.6
90° Medially 114.1±18.4 114.9±18.8 75.6±11.1 75.8±11.6
90° Laterally 114.9±17.9 114.6±17.1 75.5±11.3 75.8±10.9
180° 115.1±18.9 116.0±19.9 75.9±11.2 76.4±11.3
         
Test device: auscultatory, undersized cuff (Undercuffing Study; n=33)
 Correct 124.6±15.0 124.1±15.1 85.2±10.4 85.0±9.7
90° Medially 123.9±15.0 125.2±15.7 84.4±10.5 86.3±11.5
90° Laterally 124.5±15.6 129.5±15.4a 85.1±10.4 89.1±11.6a
180° 124.4±15.2 128.3±15.9a 84.9±10.6 89.1±11.6a
         
Test device: oscillometric, IntelliWrap cuff (Main Study; n=57)
 Correct 113.8±17.9 115.2±18.0 76.3±11.2 76.5±10.8
90° Laterally 113.8±18.2 114.9±17.1 76.3±11.7 76.1±11.8
180° 113.7±17.7 115.1±17.6 76.3±11.4 76.2±11.9
a

P<0.001 for difference between the reference and test device.