Abstract
We present a novel extended-focus optical coherence microscope (OCM) attaining 0.7 μm axial and 0.4 μm lateral resolution maintained over a depth of 40 μm, while preserving the advantages of Fourier domain OCM. Our system uses an ultra-broad spectrum from a supercontinuum laser source. As the spectrum spans from near-infrared to visible wavelengths (240 nm in bandwidth), we call the system visOCM. The combination of such a broad spectrum with a high-NA objective creates an almost isotropic 3D submicron resolution. We analyze the imaging performance of visOCM on microbead samples and demonstrate its image quality on cell cultures and ex-vivo brain tissue of both healthy and alzheimeric mice. In addition to neuronal cell bodies, fibers and plaques, visOCM imaging of brain tissue reveals fine vascular structures and sub-cellular features through its high spatial resolution. Sub-cellular structures were also observed in live cells and were further revealed through a protocol traditionally used for OCT angiography.
OCIS codes: (170.0180) Microscopy, (110.4500) Optical coherence tomography, (170.6900) Three-dimensional microscopy, (000.1430) Biology and medicine
1. Introduction
Over the past decades, optical microscopy has played a paramount role in investigating biological systems through its high spatial and temporal resolution. Confocal fluorescence microscopy [1] and light-sheet microscopy [2], through their capabilities in three-dimensional imaging, have become the mainstay for cellular and sub-cellular imaging. Nevertheless, while fluorescence provides molecular specificity, the influence of these agents on cellular processes is ambiguous, as they might interfere with the functioning of the cell. These effects combined with photobleaching ultimately hinder the possibility to perform long-term imaging.
In such studies, label-free microscopy offers an interesting alternative as it can provide wide-field images at high acquisition rates without using exogenous agents. Moreover, the absence of labels facilitates the sample preparation. Recent advances in phase microscopy [3] and ptychography [4] have allowed performing three-dimensional imaging of cellular cultures or embryos but remain limited to thin single layer structures.
Optical coherence tomography (OCT) is an interferometric imaging technique sensitive to refractive index contrast in the sample [5]. In OCT, the axial resolution is defined by the width of the illumination spectrum and an entire depth profile can be obtained from a single recording of the output spectrum. As such, only a two-dimensional scan is required to obtain a three-dimensional image.
Optical coherence microscopy (OCM), the microscopy analogue to OCT, uses high-NA objectives to obtain a higher lateral resolution. In standard OCM systems, however, the axial field of view is dictated by the Rayleigh range and thus decreases quadratically (∝ 1/ NA2) with the improvement in lateral resolution (∝ 1/ NA). This compromise can be circumvented by engineering an extended-focus illumination through the use of so-called diffraction-less beams such as Bessel beams [6]. To maintain a good collection efficiency of the scattered light signal, a Gaussian detection mode is used. Therefore, separate illumination and detection modes are required: a Bessel illumination mode and a Gaussian detection mode. This split between modes can further be exploited to filter specular reflections and obtain a dark-field OCM system [7]. The dark-field property is particularly important when investigating weakly scattering structures, such as cell samples, as it suppresses light reflected from the sample support which would otherwise strongly reduce the usable dynamic range of the detector. As such, all available dynamic range can be devoted to the desired, but weak, scattered light signal.
In this paper, we present visible spectrum optical coherence microscopy (visOCM). The system builds upon our previous dark-field OCM design, and improves its imaging capabilities for sub-cellular structures by using a large bandwidth illumination spectrum spanning visible to near-infrared wavelengths and a high-NA objective. The resulting system possesses an almost isotropic submicron resolution (0.4 μm laterally and 0.7 μm axially) maintained over a >40 μm depth of field. A conventional Fourier domain OCM (FDOCM) system without extended focus would have a much shorter depth of field of ∼ 2 μm for the same 0.4 μm lateral resolution. Hence visOCM extends the capabilities of our previous non-imaging visible light optical coherence correlation spectroscopy (OCCS) system and is optimized for three-dimensional cellular tomography [8]. Whereas recent studies have used OCT in the visible spectrum for spectroscopy [9–11], here we exploit the shorter wavelength of visible light, in conjunction with a very broad spectrum, to obtain a very high lateral and axial resolution, which is necessary for resolving sub-cellular structures. We present a strategy for dispersion compensation and the system’s 3D resolution on microbead samples. We demonstrate visOCM’s image quality and contrast on living cell cultures as well as fixed brain slices of healthy and alzheimeric mice.
Besides imaging the structure of a sample at a given time-point, there is also much interest in monitoring intracellular dynamics to understanding cell function. As such, several optical microscopy techniques have been developed to analyse cell trafficking and intracellular motility [12–14]. Recently, OCT methods developed to obtain qualitative and quantitative information on vascular function have been used to reveal sub-cellular compartments and quantify their activity [15,16]. Being a Fourier-domain method, visOCM is capable of rapidly acquiring tomograms, and therefore these dynamic signal imaging methods can be applied to visOCM as well. In this context, we demonstrate dynamic signal imaging with visOCM on living cells using a protocol used in OCT angiography.
2. Materials and methods
2.1. Optical setup
As illustrated in Fig. 1, the optical setup is based on a Mach-Zehnder interferometer, allowing for a separation of the illumination and detection modes, necessary to obtain the desired Bessel-Gauss configuration. The output of a supercontinuum laser (Koheras SuperK Extreme, NKT Photonics) is first filtered through three broadband dielectric mirrors (BB1-EO2, Thorlabs) to reject the source’s strong infrared emission. Then, the light is split by a polarizing beam-splitter (PBS251, Thorlabs), and injected in the interferometer using a single-mode fiber (P3-460B-FC-2, Thorlabs). As shown in Fig. 3(a), the spectrum in the interferometer is centred at 647 nm and has a 246 nm FWHM. After collimation, the light passes through a polarizer and then a first beamsplitter (BS1) splits the light into the reference and illumination paths, where an axicon lens (Asphericon, apex angle 176°) generates a Bessel beam. The beam is then Fourier filtered in a telescope to remove stray light originating from the axicon’s tip, after which it is steered to the scan-unit, and focused on the sample by a 40× high NA objective (Olympus, effective NA = 0.76). The back-scattered light is collected by the objective, de-scanned by the scan-unit and directed to the detection arm through a second beam splitter (BS2) where it is coupled to a custom-made spectrometer with a single-mode fiber (P3-460B-FC-2, Thorlabs). The spectrometer is comprised of a transmission grating (600 lines/mm, Wasatch Photonics) and a fast line-scan camera (Basler spL2048-140km). Dispersion matching between both interferometer arms was performed by adding a combination of prism pairs in the reference arm. The detailed dispersion matching strategy is presented in 2.2.
To obtain a dark-field configuration, the specular reflection was suppressed by spatially filtering the reflected Bessel ring through a mask in the detection arm (mask 2) and by properly filtering the stray light from the tip of the axicon lens (mask 1).
2.2. Dispersion compensation strategies
The use of a broad illumination spectrum in visOCM renders dispersion matching more challenging, particularly in the visible spectrum where the relationship between the refractive index and the wavelength becomes increasingly non-linear at shorter wavelengths. Here we opted for a physical dispersion compensation scheme where we first estimated the dispersion caused by the illumination and sample arms prior to constructing the optical system. This allowed us to calculate the best composition of prisms to place in the reference arm (Fig. 2(a)). In a second step, during the construction of the setup, we developed a real-time interface in LabView (National Instruments) to display the amount of residual dispersion and used a motorized stage to finely adjust the thickness of the different glasses (Fig. 2(b–c)).
2.2.1. Estimation of the dispersion in the optical system
Prior to constructing the optical system, we first estimated the thickness of each glass type present in the different arms of the interferometer. The glass type and respective thickness of each optical element (lenses and beamsplitters) were obtained from the data-sheets provided with each element by Thorlabs. The objective was not modelled in this initial assessment as its composition is unavailable. As depicted in Fig. 2(a) and described in Eq. (1), we then proceeded to estimate the k-dependent optical path length of each part of the system (sample arm, illumination arm and reference arm) by using the thickness of each glass and their respective dispersion curves (Sellmeier coefficients from the SCHOTT database available on https://www.refractiveindex.info) to obtain the optical path difference (OPD).
(1) |
The obtained OPD was then fit to a set of dispersion curves through a multivariate linear regression to find the thickness of the set of glasses to best balance the dispersion. As described in Eq. (2), the multivariate linear regression tries to model the OPD as the sum of the refractive indexes (in function of k) of each glass multiplied by their respective thicknesses and an error term ∊(k). By performing this analysis, we can therefore retrieve the thicknesses of each glass required to compensate the OPD of the optical system. An initial analysis of several sets of glasses revealed that combining BK7 and SF10 fully balanced the dispersion of the system.
(2) |
2.2.2. Fine matching of the dispersion
During the construction and alignment of the system, the dispersion was finely balanced by calculating the residual dispersion in the system and by adjusting the thickness of each individual glass through a pair of motorized mechanical stages. The residual dispersion was measured by placing a sample consisting of a cover glass and a mirror. Then the interferogram’s phase was calculated by Hilbert transformation [17], and a linear fit was subtracted (Fig. 2(c)). If the dispersion is perfectly balanced between the arms of the interferometer then the residual dispersion curve should be null throughout the spectrum. The motorized mechanical stages, as depicted in Fig. 2(b), consists of a stepper motor operating a cogwheel mechanism. The angle between two glass windows can be varied to fine-tune the amount of each glass type in the reference arm. Moreover, as the system is symmetrical, changes in the beam’s transverse position caused by the passage of the light through the first window are compensated by the second window. As such, varying the amount of glass causes minimal changes in the alignment of the optical system. The combination of these two effects (fine-tuning and minimal misalignment) facilitates the dispersion compensation procedure. By observing the residual dispersion and iteratively changing the thickness of each glass we could balance the dispersion of the two arms of the microscope. We used SF6 windows, instead of SF10, for dispersion fine tuning.
2.3. Image acquisition and processing
All images were acquired at a 20 kHz A-scan rate (with an integration time of 43 μs) and with a power varying from 1.5 mW to 3 mW in the back-focal plane of the objective. With the exception of the dynamic signal imaging protocol (presented in 3.4), the size of each image was 512 × 512 × 2048 pixels (x,y and k respectively). With the aforementioned parameters, the acquisition time for a single tile is ∼ 14 seconds. Large field-of-views were obtained by stitching several tomograms (each having a lateral field-of-view of either 60 μm × 60 μm, or 120 μm × 120 μm) with a 30% overlap between each tile of the mosaic in both directions. Any tilt (angle with respect to the optical axis), was corrected on both axes (x–z and y–z) prior to stitching. The tomogram processing, tilt-correction and stitching were performed through a custom-coded MATLAB (Mathworks) graphical interface. The tomograms presented in Fig. 4–7 are displayed with the intensity in logarithmic scale for visualisation purposes. The large volume presented here contains 170 tiles, which requires ∼ 40 minutes of acquisition time. The tomograms were convolved with a 3D Gaussian kernel (σx,y = 0.187 μm, σz = 0.22 μm) and were then resized to obtain an isotropic sampling using ImageJ.
2.4. Sample preparation
2.4.1. Mice brain slices
All experiments were carried out in accordance to the Swiss legislation on animal experimentation (LPA and OPAn). The protocols (VD 3056 and VD3058) were approved by the cantonal veterinary authority of the canton de Vaud, Switzerland (SCAV, Département de la sécurité et de l’environnement, Service de la consommation et des affaires vétérinaires) based on the recommendations issued by the regional ethical committee (i.e. the State Committee for animal experiments of canton de Vaud) and are inline with the 3Rs and follow the ARRIVE guidelines. Brain slices were obtained by perfusing transcardially B6SJL/f1 mice with PBS followed by 10% Formalin (HT501128, Sigma-Aldrich). The mice were injected subcutaneously with Temgesic prior to the perfusion with heparinized PBS. The extracted brains were then left in 4% PFA overnight, and then placed in a solution of 30% glucose. Finally, the brains were cut into slices using a microtome at a thickness of ∼30 μm and placed on a glass coverslide. Brain slices from 5xFAD mice, a mice model of amyloid pathology, were obtained using the same protocol. The amyloid plaques were stained using a solution of Methoxy-X04 in DMSO, which was administered through two I.P. injections 24h and 2h before the perfusion, as described by Järhling et al. [18].
2.4.2. Macrophages preparation
In addition to mice brain slice imaging, we imaged live murine macrophages (cell line RAW 264.7) with the visOCM platform. The RAW 264.7 cells were cultured in an incubator at 37°C and 5% CO2 using DMEM high glucose with pyruvate (4.5 g l-1 glucose, Roti®-CELL DMEM, Roth) supplemented with 10% fetal bovine serum and 1× penicillin-streptomycin (both gibco®, Thermo Fisher Scientific). Prior to imaging (1–2 days), the cells were seeded in FluoroDish Sterile Culture Dishes (35 mm, World Precision Instruments).
3. Results
3.1. System characterization
The system’s lateral resolution was characterized by imaging a solution of nanoparticles of 30 nm in diameter embedded in a slab of PDMS. The size of these particles is small enough to interrogate the point-spread function (PSF) of the optical system. The depth-dependence of the lateral resolution was assessed by isolating and averaging multiple (∼ 10) measurements of the PSF at 7 different depths. The lateral profile of the measured PSF at each depth was then extracted, and the position of the first zero served as a measure of the lateral resolution. As shown in Fig. 3(c) the width of the central lobe is maintained at 400 nm over 40 μm.
The axial resolution of the system was measured by imaging a mirror placed on a glass coverslip in the sample arm. The dark-field diaphragm (mask 2 in Fig. 1) was slightly opened to collect the specular reflection from the mirror. The reference power was adjusted to obtain maximum visibility of the interference pattern. The axial PSF, measured as mentioned previously, is plotted in Fig. 3(b) and has a FWHM of 0.92 μm in air, corresponding to a width of 0.69 μm in water.
3.2. Brain slice imaging
The imaging performance of visOCM was first demonstrated by imaging cortical structures in fixed mice brain slices (∼30 μm thick) of both healthy and alzheimeric mice.
The lateral and axial resolution and contrast offered by the visible spectrum allows resolving several different entities present in the brain such as fibers, vascular structures, cell bodies and amyloid plaques.
3.2.1. Neural fibers
Similarly to previous studies performed with OCM at longer wavelengths [19–22] and in full-field OCT [23], neural fibers appear as bright linear structures through their increased back-scattering. In Fig. 4(a) and (c), fibers emerge from the corpus callosum and the cingulum bundle and spread throughout the cortical column. The corpus callosum, shown in Fig. 5(d), contains a high density of fibers and can be distinguished as a bright region with orientated stripes. The cortex is characterized by a lower density of fibers with a higher variability in their orientation.
3.2.2. Vascular structures
Vascular compartments, from large calibres (penetrating arterioles and ascending venules) to the smallest capillaries can be discriminated from the background tissue as either hollow tubes (from the empty lumen) or as thin bright structures. Figure 5(c) shows the en-face and orthogonal slice of a large penetrating vessel where one can distinguish the hollow lumen from the tissue and visualise bifurcations along the propagation of the vessel ( Visualization 2 (3.5MB, AVI) ). Additionally, the edges of the vessel exhibit an increased signal, which could either be caused by a change in the scattering properties of the vessel’s membrane or its surrounding tissue (for example vascular smooth muscle). Smaller vessels, such as capillaries, can also be observed in the tomograms and appear either as dark or bright structures compared to the surrounding tissue. Figures 4(e) and (f) show both dark and bright capillary structures and show that it is possible to trace their trajectory regardless of their contrast. Similarly to the large vessel in 5(c), the dark contrast is indicative of a lack of scatterers within the lumen of the capillary. The bright contrast, on the other hand, could originate from scatterers filling the vessel’s lumen (i.e. clogging during the perfusion procedure) or from the different scattering properties of the vessel’s boundary.
3.2.3. Cell bodies
In addition to neuronal fibers and capillaries, visOCM imaging reveals different cell body types through their different contrast with respect to the extracellular space. Figures 4(b) and 5(b) present mosaics covering parts of the posterior parietal association areas (PTLp1) and the retrosplineal area (RSPd) respectively, where one can observe two main cell body types with different contrasts and shapes. As highlighted in Fig. 4(d), some of the cell bodies appear as dark spherical shapes, due to a decreased back-scattering. The second type of cell bodies visible in Fig. 4(c) and Visualization 1 (8.5MB, AVI) have similar contrast than the extracellular space and are larger. In addition to their different shapes and back-scattering properties, the two cell body types also appear to be present in different regions of the cortical column: the darker and smaller cell bodies are denser in the upper layers whereas the larger cell bodies seem more prominent in the deeper layers, closest to the corpus callosum. Cell bodies in the cornu ammonis area 1 (CA1), as shown in Fig. 6(b), are also characterized by a darker contrast, similar to the cell bodies in the upmost layers of PTLp1. Additionally, one can notice in the RSPd brain slice the presence of a small darker substructure within the body of certain cells. Figure 5(e–h) displays a selection of these cells and their dark sub-cellular structure. Although a more complete study is necessary to elucidate the nature of this feature, our experience in live-cell imaging (results displayed in Fig. 7) has shown that a similar contrast is present in what seems to be the nucleus. The orthogonal views and tile (g) of Fig. 4 show that the signal acquired with visOCM extends throughout the depth of the tissue slice, although a loss in intensity and blurring can be observed in the deeper layers.
3.2.4. Amyloid plaques
Previous work from our group has shown that amyloid plaques can be distinguished from cerebral tissue using xfOCM operating at 800 nm through their increased scattering [19]. In continuation of this work, we imaged brain slices of an alzheimeric mouse model with our novel visOCM system, with the expectation that the increased spatial resolution and different illumination wavelength would shed light on the details of these aggregates. We therefore imaged a part of the PTLp cortex and subcortical structures (CA1) of a 5xFAD mouse. As shown in Fig. 6(b) and similarly to the results in Fig. 4(b) and 5(b), the cortex is characterized by a high density of fibers and cell bodies. The subcortical region, below the corpus callosum, has a slightly lower intensity compared to the cortex and presents a line of cell bodies (CA1). Similarly to the results obtained at 800 nm [19], the amyloid plaques manifest themselves as high intensity regions with a darker core. The increased spatial resolution of the system reveals with great detail the irregular shape of these aggregates, as shown in 6(c–d). As expected, the plaques are present in cortical (see Visualization 3 (10.2MB, AVI) ) and also subcortical regions [24], where the slightly decreased intensity of the cerebral tissue provides a higher contrast between the plaques and the background. The location of the plaques was colocalized with fluorescence imaging of Methoxy-X04 using a commercial widefield microscope (Axiovert 200M, Zeiss), a 20× / 0.5 NA objective and the DAPI filter set (Excitation filter: 365 nm, dichroic mirror: 395 nm, Emission filter: 445/50 nm). As shown the overlay in tiles (c–d) of Fig. 6, the locations of the aggregates in visOCM are in agreement with the location of the labelled structures in the fluorescence image (the visOCM and fluorescence are displayed in green and red respectively). Some mismatches between both channels in the overlay are caused by the presence of labelled plaques in a different plane than in the en-face visOCM image (i.e. the fluorescence image was obtained in a conventional fluorescence microscope and does not possess any axial sectioning).
3.3. Cell imaging
In addition to the imaging of tissue structures, we performed imaging of live macrophages in a cell culture. A mosaic of tomograms of these macrophages is shown in Fig. 7(a). The capabilities of the extended-focus can be appreciated in the orthogonal views of the tomogram, where the signal extends sufficiently in depth to reveal the three-dimensional organisation of the culture, with certain cells lying on top of other cells. The strongly scattering cytoplasm of the cell (pointed in pink in Fig. 7(a)) appears as a bright structure surrounding a darker subvolume, which most likely corresponds to the cell nuclei (pointed in green). The increased lateral resolution of the system allows resolving the filopodia on certain cells (pointed in red in Fig. 7(a)).
3.4. Dynamic signal imaging
In a second step, we analysed the dynamic properties of the scattering signal from living cells. In contrast to previous dynamic signal imaging techniques using the autocorrelation function or the standard deviation of the OCT signal [13, 15, 16], we extracted the dynamic component of the back-scattering through a point-wise subtraction of the complex OCM signal (Fig. 7(b)), as developed by Srinivasan et al. for OCT angiography [25]. A time-series of scattering fluctuations per voxel was obtained by sampling each transverse position (B-scan) 32 times with a timestep of Δt = 27 ms. Each timepoint was then temporally high-pass filtered (through a point-wise complex subtraction) and then averaged. A temporally averaged image was also obtained by averaging the repeated acquisitions. The results of this operation are shown in Fig. 7(c) and Fig. 7(d) showing the averaged and dynamic signals respectively. Interestingly, in addition to the lower intensity of the cell nucleus already present in the static imaging, the dynamic imaging further reveals details within the cell body. As shown in Fig. 7(d), darker regions are present within the nuclei of the cells and brighter spots can be observed within the cytoplasm. Finally, the interface between the nuclei and the cytoplasm appears as a fine dark structure in certain cells. These differences are further revealed in Fig. 7(e) and (f) showing close-ups of a selected cell in the temporally averaged and dynamic signal image respectively. The cytoplasm and the nucleus are pointed with pink and green arrows respectively in Fig. 7(f).
4. Conclusion
In this work, we presented a novel OCM system, called visOCM, combining an extended-focus [6], a dark-field detection [7], a high-NA objective and an ultra-broad illumination spanning from the visible to the near-infrared wavelength range. As demonstrated here, the combination of these features provides an almost isotropic submicron resolution, maintained over > 40 μm in depth. The resulting system attains a lateral and axial resolution of 0.4 μm and 0.69 μm in water respectively. These characteristics were obtained by combining both a high-NA objective and a large spectrum centred in the visible wavelength range. A shorter central wavelength requires less spectral bandwidth to obtain a high axial resolution. Ultimately, we obtained an axial resolution of 0.69 μm through a spectrum with a FWHM of 246 nm. At longer central wavelengths, the bandwidth required to obtain the same optical sectioning becomes increasingly large through the quadratic dependency between the axial resolution and the central wavelength. For example, at 1300 nm, the bandwidth of the spectrum would need to be >800 nm. Such large spectrums would be limited by the spectral response and the spatial extent of current line cameras and by the bandwidth of optical elements (fibers, gratings, objectives...).
The visOCM system possesses the advantages of FDOCM systems combined with an extended-focus. As such, it has a higher SNR than time-domain based techniques (such as full-field OCM) [26] and can sample signals over a large range of timescales, from microseconds to minutes, necessary to study tissue and cell dynamics [12, 13, 15]. Moreover, scanning Fourier domain system offer more flexibility as the scanner can be programmed to image a restricted field-of-view at high speed. Through its extended focus, the visOCM system can image over a depth of 40 μm in a single acquisition (A-scan) and can perform volumetric imaging at higher speeds than standard scanning FDOCM systems, which typically require translating the optical focus in depth to obtain a 3D image [20]. Indeed, without an extended focus, the depth-of-field that an FDOCM system can acquire in a single A-scan would be limited to ∼ 2 μm (for an equal lateral resolution of 0.4 μm). Therefore, ∼ 15 2D beam scans acquired at different depths would need to be stitched together to image the same 40 μm thick volume, requiring significantly more acquisition time.
The capabilities of visOCM were demonstrated by imaging brain tissue slices of healthy and alzheimeric mice as well as macrophages cell cultures. The imaging of brain tissue with visOCM reveals several cortical structures as vessels, capillaries and cell bodies. Interestingly, these different structures exhibited a wide range of contrasts, even within the same structure type. Capillaries could be observed as both dark or bright hollow structures. Although one cannot discard changes in tissue caused by the sample preparation (i.e. clotting and vessel collapsing due to the perfusion), the bright contrast could emanate from the presence of tissue bordering the vessels (such as smooth muscle cells or pericytes). Conversely, the dark contrast arises from a lack of scatterers from the hollow lumen of the vessel. Previous studies involving OCT and OCM imaging of brain tissue showed that certain cell bodies could be identified through their different contrast within the cerebral tissue [15, 20, 22, 23, 27]. Srinivasan et al. and Tamborski et al. observed that certain neuron types could be discriminated from the tissue by their dark contrast using an OCM operating at 1300 nm and 800 nm respectively [20, 27]. By exploiting a higher lateral resolution and by shifting the illumination spectrum to the visible wavelength range, we show that this intrinsic contrast seems to vary between cell bodies. Furthermore, the increased resolution of the visOCM system reveals features previously unobserved using other OCM systems such as sub-cellular structures (potentially the cell nucleus), thin bright boundaries surrounding capillaries and their hollow lumen. Future work will focus on elucidating the causes of these different contrasts.
In a second step, we built upon a previous study performed at 800 nm by imaging brain slices of alzheimeric mice models with our novel system [19]. Similarly to our aforementioned results, the plaques appear as bright irregular structures in the tomograms obtained with visOCM. Although the exact nature of this particular contrast remains unknown, the presence of metals (such as Fe) in amyloid plaque cores, as described by Plascencia-Villa et al. [28], might provide a first hint into the cause of this phenomenon. In fact, a previous study from our group showed that the contrast of the Langerhans islets in OCM images originated from the presence of Zn crystals [29]. Alternatively, this contrast could also arise from polarization effects as the plaques have been shown to have birefringent properties [30].
Finally, we demonstrated the performance of visOCM by imaging live-cells in cultures. The extended-focus and high acquisition rates available with the platform enable fast imaging of the three-dimensional structure of cell cultures. The shift in illumination wavelength and increase in NA provides the resolution necessary to identify thin structures such as filopodias and sufficient contrast to visualize seemingly sub-cellular structures. Recent studies have highlighted the importance of studying the dynamic properties of cellular back-scattering to understand cellular function [16, 31]. In this context, we explored the possibility to perform dynamic imaging, as already performed in full-field OCM [16], FDOCM [12] and phase imaging [14], with our novel imaging platform to further discriminate between sub-cellular compartments. We applied a protocol devised for OCT angiography [25] and show that a dynamic contrast can be obtained in a culture of macrophages. Apelian et al. and Leroux et al. obtain a similar contrast using dynamic full-field OCT and the ACF and demonstrate that their contrast is caused by ATP-consuming motion (likely mitochondria or other organelles) [16, 31]. Although more work is needed to identify the different regions and the nature of the signals obtained here, it is likely that the origin of our contrast involves similar mechanisms. Moreover, further work would be necessary to compare and assess the advantages and differences of the methods for obtaining contrast based on the dynamics of the signal (i.e. auto-correlation function [13, 15, 31], standard deviation [16] and point-wise subtraction [25]). Overall, the combination of the extended-focus, the high isotropic resolution and high acquisition rates make visOCM an ideal platform to monitor fast processes occurring at the sub-cellular level in cell cultures.
In addition to our demonstration of the capabilities of our novel visOCM system, we have introduced a strategy for physical dispersion compensation. We used a multivariate linear regression to model the OPD of the system prior to alignment and present a processing pipeline and a hardware unit to finely minimize the dispersion during the construction of the microscope. In this work, we opted for a hardware dispersion compensation strategy, however numerical dispersion compensation techniques could also have been used and will be explored in future work.
visOCM offers label-free 3D imaging of tissue and cell structures but remains limited through its lack of specificity. Our results show the ability to discriminate between structures in tissue through their individual back-scattering properties and potentially through their dynamic signatures (as revealed by our dynamic imaging). Nevertheless, these contrast mechanisms are limited and do not always provide the desired molecular specificity. In such regards, the vi-sOCM platform could be augmented with a collinear fluorescence channel, as in multimodal OCT platforms [32–34]. This configuration would allow molecular-specific fluorescent imaging to be complemented with visOCM imaging, providing information on the overall structural context of the organism under investigation. In addition to fluorescence, the visible spectrum could be used to explore the spectroscopic signatures of specific molecules. Spectroscopic OCT has shown great promise in its ability to provide additional contrast mechanisms [9, 35]. Having an illumination in the visible wavelength range could be used to visualize and discriminate between different endogeneous and/or exogenous contrast agents within the imaged tissue or cells.
Finally, even though the visOCM system was designed for imaging cells and relatively thin layers of tissue (∼ 30 μm), thicker tissue slices can be imaged by translating the focus in depth. Placing the focus deeper in a tissue might however cause a loss of resolution through aberrations and increased dispersion, for which numerical and depth-dependent dispersion compensation techniques might be necessary. Ultimately, the imaging depth obtainable in tissue would be limited by the penetration depth of the visible illumination, measured here to be ∼ 110 μm.
Acknowledgments
We thank Kristin Grussmayer for preparing the cells and B. Deplancke (LSBG, EPFL) for kindly giving us the murine macrophages cell line RAW 264.7.
Funding
This study was partially supported by the Swiss National Science Foundation (205321L_135353 and 205320L_150191), the Commission for Technology and Innovation (13964.1 PFLS-LS and 17537.2 PFLS-LS) and by the EU Framework Programme for Research and Innovation (686271).
Disclosures
The authors declare that there are no conflicts of interest related to this article.
References and links
- 1.Pawley J., Handbook of Biological Confocal Microscopy (Springer, 2006). 10.1007/978-0-387-45524-2 [DOI] [Google Scholar]
- 2.Huisken J., Swoger J., Del Bene F., Wittbrodt J., Stelzer E. H. K., “Optical sectioning deep inside live embryos by selective plane illumination microscopy,” Science 305(5686), 1007–1009 (2004). 10.1126/science.1100035 [DOI] [PubMed] [Google Scholar]
- 3.Kim T., Zhou R., Mir M., Babacan S. D., Carney P. S., Goddard L. L., Popescu G., “White-light diffraction tomography of unlabelled live cells,” Nat. Photon. 8, 256–263 (2014). 10.1038/nphoton.2013.350 [DOI] [Google Scholar]
- 4.Tian L., Waller L., “3D intensity and phase imaging from light field measurements in an LED array microscope,” Optica 2(2), 104–111 (2015). 10.1364/OPTICA.2.000104 [DOI] [Google Scholar]
- 5.Fercher A. F., Drexler W., Hitzenberger C. K., Lasser T., “Optical coherence tomography - development, principles, applications,” Rep. Prog. Phys. 66, 239–303 (2003). 10.1088/0034-4885/66/2/204 [DOI] [Google Scholar]
- 6.Leitgeb R. A., Villiger M., Bachmann A. H., Steinmann L., Lasser T., “Extended focus depth for Fourier domain optical coherence microscopy,” Opt. Lett. 31(16), 2450–2452 (2006). 10.1364/OL.31.002450 [DOI] [PubMed] [Google Scholar]
- 7.Villiger M., Pache C., Lasser T., “Dark-field optical coherence microscopy,” Opt. Lett. 35(20), 3489–3491 (2010). 10.1364/OL.35.003489 [DOI] [PubMed] [Google Scholar]
- 8.Broillet S., Szlag D., Bouwens A., Maurizi L., Hofmann H., Lasser T., Leutenegger M., “Visible light optical coherence correlation spectroscopy,” Opt. Express 22(18), 21944–21957 (2014). 10.1364/OE.22.021944 [DOI] [PubMed] [Google Scholar]
- 9.Robles F. E., Wilson C., Grant G., Wax A., “Molecular imaging true-colour spectroscopic optical coherence tomography,” Nat. Photon. 5(12), 744–747 (2011). 10.1038/nphoton.2011.257 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Yi J., Wei Q., Liu W., Backman V., Zhang H. F., “Visible-light optical coherence tomography for retinal oximetry,” Opt. Lett. 38(11), 1796–1798 (2013). 10.1364/OL.38.001796 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Chong S. P., Merkle C. W., Leahy C., Radhakrishnan H., Srinivasan V. J., “Quantitative microvascular hemoglobin mapping using visible light spectroscopic Optical Coherence Tomography,” Biomed. Opt. Express 6(4), 1429–1450 (2015). 10.1364/BOE.6.001429 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Sison M., Chakrabortty S., Extermann J., Nahas A., Marchand P. J., Lopez A., Weil T., Lasser T., “3D time-lapse imaging and quantification of mitochondrial dynamics,” Sci. Rep. 7, 43275 (2017). 10.1038/srep43275 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Oldenburg A., Yu X., Gilliss T., Alabi O., Taylor R. M., II, Troester M. A., “Inverse-power-law behavior of cellular motility reveals stromal-epithelial cell interactions in 3D co-culture by OCT fluctuation spectroscopy,” Optica 2, 877–885 (2015). 10.1364/OPTICA.2.000877 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Ma L., Rajshekhar G., Wang R., Bhaduri B., Sridharan S., Mir M., Chakraborty A., Iyer R., Prasanth S., Millet L., Gillette M. U., Popescu G., “Phase correlation imaging of unlabeled cell dynamics,” Sci. Rep. 6, 32702 (2016). 10.1038/srep32702 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Lee J., Radhakrishnan H., Wu W., Daneshmand A., Climov M., Ayata C., Boas D. A., “Quantitative imaging of cerebral blood flow velocity and intracellular motility using dynamic light scattering-optical coherence tomography,” J. Cereb. Blood Flow Metab. 33(6), 819–825 (2013). 10.1038/jcbfm.2013.20 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Apelian C., Harms F., Thouvenin O., Boccara A. C., “Dynamic full field optical coherence tomography: sub-cellular metabolic contrast revealed in tissues by interferometric signals temporal analysis,” Biomed. Opt. Express 7(4), 1511–1524 (2016). 10.1364/BOE.7.001511 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Wojtkowski M., Srinivasan V. J., Ko T. H., Fujimoto J. G., Kowalczyk Andrzej, Duker J. S., “Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation,” Opt. Express 12(11), 707–709 (2004). 10.1364/OPEX.12.002404 [DOI] [PubMed] [Google Scholar]
- 18.Jährling N., Becker K., Wegenast-Braun B. M., Grathwohl S. A., Jucker M., Dodt H.-U., “Cerebral β-amyloidosis in mice investigated by ultramicroscopy,” PLOS ONE 10(5), 1–13 (2015). 10.1371/journal.pone.0125418 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Bolmont T., Bouwens A., Pache C., Dimitrov M., Berclaz C., Villiger M., Wegenast-Braun B. M., Lasser T., Fraering P. C., “Label-free imaging of cerebral β-amyloidosis with extended-focus optical coherence microscopy,” J. Neurosci. 32, 14548–14556 (2012). 10.1523/JNEUROSCI.0925-12.2012 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Srinivasan V. J., Radhakrishnan H., Jiang J. Y., Barry S., Cable A. E., “Optical coherence microscopy for deep tissue imaging of the cerebral cortex with intrinsic contrast,” Opt. Express 20(3), 2220–2239 (2012). 10.1364/OE.20.002220 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Leahy C., Radhakrishnan H., Srinivasan V. J., “Volumetric imaging and quantification of cytoarchitecture and myeloarchitecture with intrinsic scattering contrast,” Biomed. Opt. Express 4(10), 1978–1990 (2013). 10.1364/BOE.4.001978 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Magnain C., Augustinack J. C., Konukoglu E., Frosch M. P., Sakadzic S., Varjabedian A., Garcia N., Wedeen V. J., Boas D. A., Fischl B., “Optical coherence tomography visualizes neurons in human entorhinal cortex,” Neurophotonics 2(1), 015004 (2015). 10.1117/1.NPh.2.1.015004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Assayag O., Grieve K., Devaux B., Harms F., Pallud J., Chretien F., Boccara C., Varlet P., “Imaging of non-tumorous and tumorous human brain tissues with full-field optical coherence tomography,” Neuroimage Clin. 2, 549–557 (2013). 10.1016/j.nicl.2013.04.005 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Jawhar S., Trawicka A., Jenneckens C., Bayer T. A., Wirths O., “Motor deficits, neuron loss, and reduced anxiety coinciding with axonal degeneration and intraneuronal A aggregation in the 5XFAD mouse model of Alzheimer’s disease,” Neurobiol. Aging 33(1), 196 (2012). 10.1016/j.neurobiolaging.2010.05.027 [DOI] [PubMed] [Google Scholar]
- 25.Srinivasan V. J., Jiang J. Y., Yaseen M. A., Radhakrishnan H., Wu W., Barry S., Cable A. E., Boas D. A., “Rapid volumetric angiography of cortical microvasculature with optical coherence tomography,” Opt. Lett. 35(1), 43–45 (2010). 10.1364/OL.35.000043 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Leitgeb R., Hitzerberger C. K., Fercher A. F., “Performance of fourier domain vs time domain optical coherence tomography,” Opt. Express 11(8), 889–894 (2003). 10.1364/OE.11.000889 [DOI] [PubMed] [Google Scholar]
- 27.Tamborski S., Lyu H. C., Dolezyczek H., Malinowska M., Wilczynski G., Szlag D., Lasser T., Wojtkowski M., Szkulmowski M., “Extended-focus optical coherence microscopy for high-resolution imaging of the murine brain,” Biomed. Opt. Express 7(11), 4400–4414 (2016). 10.1364/BOE.7.004400 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Plascencia-Villa G., Ponce A., Collingwood J. F., Arellano-Jiméanez M. J., Zhu X., Rogers J. T., Betancourt I., José-Yacamán M., Perry G., “High-resolution analytical imaging and electron holography of magnetite particles in amyloid cores of Alzheimer’s disease,” Sci. Rep. 6, 24873 (2016). 10.1038/srep24873 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Berclaz C., Schmidt-Christensen A., Szlag D., Extermann J., Hansen L., Bouwens A., Villiger M., Goulley J., Schuit F., Grapin-Botton A., Lasser T., Holmberg D., “Longitudinal three-dimensional visualisation of autoimmune diabetes by functional optical coherence imaging,” Diabetologia 59(3), 550–559 (2016). 10.1007/s00125-015-3819-x [DOI] [PubMed] [Google Scholar]
- 30.Baumann B., Woehrer A., Ricken G., Augustin M., Mitter C., Pircher M., Kovacs G. G., Hitzenberger C. K., “Visualization of neuritic plaques in Alzheimer’s disease by polarization-sensitive optical coherence microscopy,” Sci. Rep. 7, 43477 (2017). 10.1038/srep43477 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Leroux C.-E., Bertillot F., Thouvenin O., Boccara A. C., “Intracellular dynamics measurements with full field optical coherence tomography suggest hindering effect of actomyosin contractility on organelle transport,” Biomed. Opt. Express 7(11), 4501 (2016). 10.1364/BOE.7.004501 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Auksorius E., Bromberg Y., Motiejūnaitė R., Pieretti A., Liu L., Coron E., Aranda J., Goldstein A. M., Bouma B. E., Kazlauskas A., Tearney G. J., “Dual-modality fluorescence and full-field optical coherence microscopy for biomedical imaging applications,” Biomed. Opt. Express 3(3), 661–666 (2012). 10.1364/BOE.3.000661 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Makhlouf H., Perronet K., Dupuis G., Lévêque-Fort S., Dubois Arnaud, “Simultaneous optically sectioned fluorescence and optical coherence microscopy with full-field illumination,” Opt. Lett. 37(10), 1613–1615 (2012). 10.1364/OL.37.001613 [DOI] [PubMed] [Google Scholar]
- 34.Thouvenin O., Fink M., Boccara C., “Dynamic multimodal full-field optical coherence tomography and fluorescence structured illumination microscopy,” J. Biomed. Opt. 22(2), 026004 (2017). 10.1117/1.JBO.22.2.026004 [DOI] [PubMed] [Google Scholar]
- 35.Morgner U., Drexler W., Kärtner F. X., Li X. D., Pitris C., Ippen E. P., Fujimoto J. G., “Spectroscopic optical coherence tomography,” Opt. Lett. 25(2), 111–113 (2000). 10.1364/OL.25.000111 [DOI] [PubMed] [Google Scholar]