Table 4. Areas under the receiver operating characteristic curves and sensitivities at fixed specificities for the sector parameters in Spectralis and Stratus OCT.

Parameters AUROC (SE) Sensitivity/Specificity
Specificity >80.0% Specificity >95%
Spectralis OCT (sectors)
 Temporal-superior 0.885 (0.052) 62.5/81.5 41.7/96.3
 Nasal-superior 0.502 (0.094) 4.2/81.5 0/96.3
 Nasal-inferior 0.648 (0.089) 37.5/81.5 25.0/96.3
 Temporal-inferior 0.974 (0.020) 95.8/85.2 66.7/96.3
Stratus OCT (clock hours)
 12 superior 0.627 (0.098) 43.5/82.1 39.1/96.4
 11 0.842 (0.061) 73.9/82.1 56.5/96.4
 10 0.779 (0.080) 78.3/82.1 47.8/96.4
 9 temporal 0.772 (0.074) 60.9/82.1 21.7/96.6
 8 0.840 (0.060) 69.6/82.1 39.1/96.4
 7 0.918 (0.042) 91.3/82.1 73.9/96.4
 6 inferior 0.878 (0.053) 87.0/82.1 34.8/96.4
 5 0.551 (0.093) 17.4/82.1 13.0/96.4
 4 0.584 (0.090) 13.0/82.1 0/96.4
 3 nasal 0.689 (0.084) 8.7/85.7 0/96.4
 2 0.634 (0.089) 17.4/82.1 4.3/96.4
 1 0.513 (0.099) 30.4/82.1 17.4/96.4

AUROC: Area under the receiver operating characteristic curves; OCT: Optical coherence tomography; SE: Standard error.