Skip to main content
. Author manuscript; available in PMC: 2018 Jun 1.
Published in final edited form as: Int J Radiat Oncol Biol Phys. 2017 Feb 12;98(2):438–446. doi: 10.1016/j.ijrobp.2017.02.006

Table 2.

FDA Determined Causes, Radiation Oncology vs. All Other Class 2 Device Recalls

FDA Determined Cause Radiation Oncology Devices (n=502) Other Class 2 Devices (n=9,534)*
Count Percent Count Percent
Software 246 49.0 958 10.0
Device Design/Change Control 84 16.7 991 10.4
Other/Under Investigation 81 16.1 3,163 33.2
Labeling 24 4.8 675 7.1
Material/Component 21 4.2 1,138 11.9
Process 20 4.0 1,024 10.7
Component 15 3.0 357 3.7
Employee/Use Error 7 1.4 297 3.1
Radiation Control for Health and Safety Act 3 0.6 110 1.2
Equipment Maintenance 1 0.2 111 1.2
Packaging 0 0.0 507 5.3
*

For brevity, the following determined causes were omitted, as 0% of ROD recalls and <1% of other device recalls cited them: Counterfeit, Environmental control, PMA, Reprocessing Controls, Storage, Vendor change control