Skip to main content
. 2017 May 2;19(Suppl D):D190–D211. doi: 10.1093/eurheartj/sux031

Table 1.

Characteristics of frequency response that an electrocardiograph should have in relation to the various intended uses

Aim Age of patient Mode High pass filter (Hz) Low pass filter (Hz)
Diagnostics Adult Manual 0.05 150
Diagnostics Adult Automatic 0.05–0.67To be defined 150
Diagnostics Child Manual 0.05 250
Diagnostics Child Automatic 0.05–0.67To be defined 250