Table 2.
Hydrodynamic properties of oDF and hDF derived engineered tissue valve at aortic conditions.
Ovine TEHV | Human TEHV | |
---|---|---|
Systolic ΔP | 2.9 mmHg | 2.8 mmHg |
Diastolic ΔP | 94 mmHg | 83 mmHg |
Average Flow Rate | 6.6 LPM | 5.5 LPM |
Peak Flow Rate | 14.2 LPM | 10.3 LPM |
Geometric Area1 (%) | 97 % | 95 % |
Regurgitant Fraction | 1.1 % | 3.2 % |