Skip to main content
. 2017 Mar 8;6(3):e004853. doi: 10.1161/JAHA.116.004853

Table 3.

Subgroup Analysis on the Primary End Point, ie, the Proportion of Patients in Sinus Rhythm After 4 Hours

PB Waveform BTE Waveform Ratio (95% CI) P Value
Sinus rhythm after 4 hours
Atrial fibrillation patients (n=116) 35 (58) 47 (84) 1.4 (1.1–1.8) 0.002
Atrial flutter patients (n=18) 7 (77) 9 (100) 1.3 (0.9–1.9) 0.13
Patient receiving amiodarone (n=24) 2 (33) 16 (89) 2.7 (0.9–8.4) 0.05
Patients not receiving amiodarone (n=110) 40 (64) 40 (85) 1.4 (1.1–1.7) 0.01

Data are numbers of successful cardioversions (%) and estimates are reported with their corresponding 95% CI. BTE indicates biphasic truncated exponential; PB, pulsed biphasic.