Table 1.

A summary of correlation ratios for the sampling region (dose per escaped Cherenkov photon), % relative standard deviation values and sampling depths for x-ray beams, the two values in each box represent those values for entrance and then exit sides of the tissue.

Factor
Examined
Range of correlation ratio
(Dose (Gy) per escaped
Cherenkov photon×10−7)
Relative standard
deviation (%)
(std. dev./mean)
Sampling depth (mm)
Entrance Exit Entrance Exit Entrance Exit
Skin color range light to dark 3.5 ~ 8.6 1.6 ~ 4.1 42* 45* 5.7 ~ 6.2 2.1 ~ 2.7
Nominal photon energy (MV) 4–15 MV 4.3 ~ 2.5 1.9 ~ 1.1 21* 22* 5.4 ~ 6.0 2.1 ~ 2.0
Curved surface 0–180° (10cm diam.) 3.4 ~ 2.2 2.2 ~ 1.6 18* 14* 5.0 ~ 3.1 3.1 ~ 2.4
Field size0.5 to 20 cm square 3.8 ~ 3.4 1.4 ~ 1.7 4.0 5.0* 6.3 ~ 5.1 2.0 ~ 2.1
Thickness 2.520 cm 3.5 ~ 3.6 1.6 ~ 1.5 0.73 2.3 5.7 ~ 5.7 2.1 ~ 2.1
Curvature 2.520cm diam. 3.0 ~ 2.9 1.9 ~ 1.8 1.4 1.2 4.0 ~ 4.3 2.4 ~ 2.6
SSD 80–120 cm 3.5 ~ 3.6 1.5 ~ 1.6 0.21 1.1 5.7 ~ 5.7 2.1 ~ 2.1

Those values with greater than a 5% effect are noted (*)