TABLE 2.
MDCT and 3DTEE measurements of the aortic annulus area, perimeter, and derived diameters shown along with inter-technique differences (actual values in line 3 and percent of the mean measured value in line 4). None of the inter-technique differences were statistically significant
Area (mm2) | Perimeter (mm) | Area-derived diameter (mm) | Perimeter-derived diameter (mm) | |
---|---|---|---|---|
MDCT | 504±100 | 83±8 | 25±2 | 26±3 |
Semi-automated 3D analysis | 498±111 | 80±9 | 25±3 | 25±3 |
Inter-technique difference | −6.6±46.4 (P=.34) | −3.0±3.9 (P<.001) | −0.2±1.2 (P=.24) | −1.0±1.3 (P<.001) |
% Difference | −1±10% | −4±5% | −1±5% | −4±5% |
Mitral valve software | 527±111 | 82±9 | ||
Inter-technique difference | 23±62 (P=.01) | −0.4±5.5 (P=.61) | ||
% Difference | 4±12% | −1±7% | ||
3D planimetry | 463±108 | |||
Inter-technique difference | 42±66 (P<.001) | |||
% Difference | −9±15 |
Data are shown in the same format as the top portion of table, but this time for two subgroups of patients according to the phase of the cardiac cycle at which their computed tomography images were acquired.
MDCT=multi-detector comuted tomography; 3DTEE=three-dimensional transesophageal echocardiography.