Published in final edited form as: Echocardiography. 2017 Mar 27;34(5):690–699. doi: 10.1111/echo.13483

TABLE 2.

MDCT and 3DTEE measurements of the aortic annulus area, perimeter, and derived diameters shown along with inter-technique differences (actual values in line 3 and percent of the mean measured value in line 4). None of the inter-technique differences were statistically significant

Area (mm2) Perimeter (mm) Area-derived diameter (mm) Perimeter-derived diameter (mm)
MDCT  504±100    83±8    25±2    26±3
Semi-automated 3D analysis  498±111    80±9    25±3    25±3
 Inter-technique difference −6.6±46.4 (P=.34) −3.0±3.9 (P<.001) −0.2±1.2 (P=.24) −1.0±1.3 (P<.001)
 % Difference    −1±10%    −4±5%    −1±5%    −4±5%
Mitral valve software  527±111    82±9
 Inter-technique difference    23±62 (P=.01) −0.4±5.5 (P=.61)
 % Difference      4±12%    −1±7%
3D planimetry  463±108
 Inter-technique difference    42±66 (P<.001)
 % Difference    −9±15

Data are shown in the same format as the top portion of table, but this time for two subgroups of patients according to the phase of the cardiac cycle at which their computed tomography images were acquired.

MDCT=multi-detector comuted tomography; 3DTEE=three-dimensional transesophageal echocardiography.