Published in final edited form as: Echocardiography. 2017 Mar 27;34(5):690–699. doi: 10.1111/echo.13483

TABLE 4.

Results of reproducibility analysis of MDCT and 3DTEE measurements of the aortic annulus area, perimeter, and derived diameters obtained in a subgroup of 10 study patients

Area (mm2) Perimeter (mm) Area-derived diameter (mm) Perimeter-derived diameter (mm)
MDCT
 Inter-observervariability (%) [and ICC] 4.6±2.5 [.83] 3.2±1.6 [.74] 2.3±1.3 [.83] 3.2±1.6 [.74]
 Intra-observer variability (%) [and ICC] 4.6±3.5 [.80] 1.9±2.1 [.83] 2.3±1.7 [.80] 1.9±2.1 [.83]
Semi-automated 3D analysis
 Inter-observer variability (%) [and ICC] 8.4±4.5 [.70] 4.2±2.2 [.72] 4.1±2.2 [.72] 4.2±2.1 [.71]
 Intra-observer variability (%) [and ICC] 4.3±5.6 [.83] 2.2±2.5 [.85] 2.1±2.9 [.81] 2.2±2.5 [.85]
Mitral valve software
 Inter-observer variability (%) [and ICC] 10.5±7.0 [.64] 5.7±3.5 [.61]
 Intra-observer variability (%) [and ICC] 13.8±9.4 [.31] 7.8±6.8 [.38]
3D planimetry
 Inter-observer variability (%) [and ICC] 8.0±6.6 [.80]
 Intra-observer variability (%) [and ICC] 5.7±3.4 [.89]

Data for inter- and intra-observer variability are shown as absolute differences between pairs of repeated measurements in percent of their mean value.

MDCT=multi-detector comuted tomography; 3DTEE=three-dimensional transesophageal echocardiography; ICC=intra-class correlation coefficient.