Table 3. Sensitivity and specificity at different cutoff values.

The cutoff values are expressed in Reader Units (RUs). Sensitivity and specificity values are reported in %, with the 95% confidence intervals are reported in brackets. Ov = O. volvulus, Wb = W. bancrofti, Mp = M. perstans, Ss = S. stercoralis, EN = Endemic Normals, NEN = Non-Endemic Normals.

Cutoff Visual 100 RUs 200 RUs 400 RUs 600 RUs
Sensitivity Loa (%) 94 (88–97) 88 (81–93) 84 (76–90) 80 (71–86) 71 (62–78)
Specificity Ov (%) 82 (73–88) 83 (74–89) 89 (10–94) 94 (87–97) 96 (90–98)
Specificity Wb (%) 87 (75–94) 84 (71–92) 90 (78–96) 98 (89–100) 100 (93–100)
Specificity Mp (%) 88 (64–98) 88 (64–98) 94 (72–100) 100 (91–100) 100 (91–100)
Specificity Ss (%) 100 (91–100) 100 (91–100) 100 (91–100) 100 (91–100) 100 (91–100)
Specificity EN and NEN (%) 100 (95–100) 100 (95–100) 100 (95–100) 100 (95–100) 100 (95–100)