Skip to main content
. 2017 Jul 27;12(7):e0181958. doi: 10.1371/journal.pone.0181958

Table 2. Comparison of flatness and symmetry for the results obtained using the EBT3 film and ionization chamber.

EBT3 film Ionization chamber
Flatness (%) Symmetry (%) Flatness (%) Symmetry (%)
3 × 3 cm2 8.30 0.63 8.90 0.83
5 × 5 cm2 3.86 0.76 4.90 0.63
10 × 10 cm2 2.05 0.65 1.83 0.49
15 × 15 cm2 1.96 0.71 1.71 0.75