Table 2.
Fetal DNA fraction, % | No. of positive droplets | Call rate, %* | Accuracy, %† |
---|---|---|---|
5 | 200 | 0.7 | 84.8 |
400 | 6.5 | 93.5 | |
600 | 16.1 | 93.2 | |
800 | 24.4 | 95.4 | |
1000 | 33.5 | 94.8 | |
1200 | 40.3 | 96.0 | |
1400 | 47.3 | 95.9 | |
1600 | 52.1 | 96.2 | |
1800 | 57.2 | 96.7 | |
2000 | 60.7 | 97.1 | |
10 | 200 | 28.8 | 93.6 |
400 | 56.9 | 95.8 | |
600 | 72.2 | 97.2 | |
800 | 80.0 | 98.1 | |
1000 | 86.9 | 98.6 | |
1200 | 90.8 | 99.0 | |
1400 | 93.5 | 99.4 | |
1600 | 95.6 | 99.5 | |
1800 | 96.5 | 99.6 | |
2000 | 97.5 | 99.8 | |
15 | 200 | 60.9 | 96.9 |
400 | 85.1 | 98.5 | |
600 | 93.2 | 99.4 | |
800 | 97.4 | 99.7 | |
1000 | 98.3 | 99.9 | |
1200 | 99.4 | 99.9 | |
1400 | 99.7 | 100.0 | |
1600 | 99.9 | 100.0 | |
1800 | 100.0 | 100.0 | |
2000 | 100.0 | 100.0 | |
20 | 200 | 84.1 | 98.2 |
400 | 96.4 | 99.7 | |
600 | 99.2 | 99.9 | |
800 | 99.8 | 100.0 | |
1000 | 99.9 | 100.0 | |
1200 | 100.0 | 100.0 | |
1400 | 100.0 | 100.0 | |
1600 | 100.0 | 100.0 | |
1800 | 100.0 | 100.0 | |
2000 | 100.0 | 100.0 | |
30 | 200 | 98.6 | 99.8 |
400 | 100.0 | 100.0 | |
600 | 100.0 | 100.0 | |
800 | 100.0 | 100.0 | |
1000 | 100.0 | 100.0 | |
1200 | 100.0 | 100.0 | |
1400 | 100.0 | 100.0 | |
1600 | 100.0 | 100.0 | |
1800 | 100.0 | 100.0 | |
2000 | 100.0 | 100.0 |
Call rate (%) corresponds to the percentage of samples that achieves the SPRT threshold to make a call.
Accuracy (%) refers to the percentage of samples showing a correct call.