Skip to main content
. 2017 Jul 28;7:6764. doi: 10.1038/s41598-017-07119-6

Figure 1.

Figure 1

Schematic representation of the dual lactate-glucose biosensor developed from ceramic-based multisite microelectrode arrays (MEA) (125 µm thick) containing 4 platinum recording sites in line (R1, 50 × 150 µm2, spacing 50 µm). Sites 1 and 3 sites (active sites) were coated with a cocktail solution containing Lactate Oxidase (LOx) or Glucose Oxidase (GOx), BSA and glutaraldehyde (GA). Sites 2 and 4 (sentinel sites) were coated with the inactive protein matrix. The sites were further modified with an exclusion layer of meta-phenylenediamine (m-PD), to improve selectivity, and a diffusional barrier of polyurethane (PU), to extend the linear range for the substrate detection.