Intra- and inter-day precision and accuracy. Intra- and inter-assay precision, expressed as coefficient of variation (%CV), and recovery data evaluated at each Quality Control (QC) level. Standard deviation (SD).

Intra-Assay Inter-Assay Recovery %
Compound QC Level Mean Response SD %CV Mean Response SD %CV Mean SD
CORT QC1 0.13 0.02 13.47 0.14 0.01 8.43 50.12 13.18
QC2 0.52 0.01 2.35 0.52 0.03 5.12 33.12 0.73
QC3 2.88 0.15 5.09 2.57 0.13 5.14 44.29 1.98
11-DECOL QC1 0.27 0.02 7.97 0.15 0.02 12.41 46.99 6.64
QC2 0.67 0.01 1.63 0.72 0.06 7.83 67.45 6.67
QC3 2.19 0.11 4.98 2.26 0.21 9.48 58.18 2.83
ADIONE QC1 0.83 0.04 4.99 0.99 0.03 3.15 23.26 7.16
QC2 2.14 0.10 4.77 2.30 0.15 6.72 22.01 2.26
QC3 11.99 0.13 1.12 11.85 0.32 2.73 24.53 2.23
TESTO QC1 0.67 0.09 13.82 0.56 0.05 8.77 19.50 2.92
QC2 2.14 0.31 14.67 1.92 0.09 4.51 37.46 0.48
QC3 6.39 0.73 11.49 5.29 0.74 13.94 44.85 5.91
17-OHP QC1 0.21 0.03 15.24 0.35 0.04 10.84 22.78 9.92
QC2 0.78 0.07 8.86 1.04 0.06 5.99 30.61 6.49
QC3 6.25 0.17 2.77 6.76 0.62 9.15 50.64 2.60
PROG QC1 0.83 0.03 3.20 1.01 0.07 6.90 39.56 4.31
QC2 2.26 0.21 9.54 2.95 0.19 6.70 30.60 1.92
QC3 13.92 0.10 0.72 14.77 1.22 8.30 40.43 7.50