Skip to main content
. 2017 Jun;7(3):310–317. doi: 10.21037/qims.2017.05.03

Table 7. Urinary bladder doses based on ICRP 106.

Patient Calculated Dose (mGy) Measured Dose (mGy) Percentage difference (%)
1 34.10 0.64 98.12
2 36.70 0.31 99.16
3 36.51 0.25 99.32
4 41.46 0.66 98.41
5 36.17 3.63 89.96
m ± SD 36.99± 2.71 1.10±1.43 96.99
6 36.27 1.78 95.09
7 42.14 1.06 97.48
8 39.92 0.29 99.27
9 38.87 1.06 97.27
m ± SD 39.30±2.44 1.05±0.61 97.28

m, mean; SD, standard deviation; LK, left kidney; RK, right kidney.