. 2017 Aug 1;6:103. doi: 10.1186/s40249-017-0314-2

Table 3.

The performance of the pLDH ELISA, malaria antibody ELISA, and the RDT for the detection of malaria parasites, as compared with the microscopic method (GM)

Parameter pLDH ELISA % (CI) Antibody ELISA % (CI) RDT % (CI)
Sensitivity 86.0 (77.4 – 92.8) 69.9 (60.1 – 78.6) 88.0 (80.0 – 94.0)
Specificity 92.7 (85.6 – 97.0) 80.3 (69.9 – 88.3) 99.1 (98.0 – 99.3)
PPV 91.6 (83.4 – 96.5) 81.8 (72.2 – 89.2) 89.8 (78.0 – 92.3)
NPV 88.1 (80.2 – 93.7) 67.7 (57.4 – 76.9) 99.0 (98.2 – 99.5)
FPR 7.3 (5.8 – 9.1) 19.7 (17.3 – 22.3) 1.2 (0.6 – 2.1)
FNR 2.3 (1.5 – 3.4) 30.1 (27.3 – 33.1) 12 (6.4 – 20.0)
Agreement between tests 89.7 (84.4 – 93.7) 74.5 (67.5 – 80.6) 97.9 (96.9 – 98.6)

Sensitivity = [true positive/(true positive + false negative) × 100]; specificity = [true negative/(true negative + false positive) × 100; PPV = [true positive/(true positive + false positive) × 100]; NPV = [true negative/(true negative + false negative) × 100]; Agreement = [true positive + true negative/N × 100]; FPR = 1 ˗ specificity; FNR = 1 ˗ sensitivity