Skip to main content
. 2017 Apr 1;33(9):1313–1322. doi: 10.1007/s10554-017-1124-0

Fig. 2.

Fig. 2

Rectangular shaped struts of Absorb BVS (A1) was automatically detected by QCU-CMS (v.14.9) after automatic detection of luminal contour (red contour) (A2) and interpolated luminal contour (blue contour) (A3), protrusion/embedment distances were detected using the methodology described by Sotomi et al. [20] (A3). The illustration (A4) demonstrates the protrusion distance (a), embedment depth (b), interpolated luminal contour (c) and embedment line (d). The adluminal surface of the circular struts of Mirage BRMS (B1) were detected during automatic detection of luminal contour (red) (B2). After automatic detection of interpolated luminal contour (blue contour) (B3) protrusion distances were measured automatically by the software (B3). The illustration (B4) demonstrates the protrusion distance (a), embedment depth (b), interpolated luminal contour (c) and embedment line (d)