Table 2.

Overview of spatio-temporal gait parameters for eleven healthy subjects and four PD patients (n = 1166 steps). Shown are the mean parameters (SD), Pearson correlation coefficient r, bias (SD), absolute error (SD) and the relative absolute error.

Parameter Sensor Camera r Bias Abs. Error Abs. Error (%)
Stride time (s) 1.15 (0.18) 1.15 (0.18) 1.00 −0.000 (0.016) 0.013 (0.010) 1.1
Stance time (s) 0.74 (0.14) 0.70 (0.13) 0.99 0.037 (0.020) 0.037 (0.019) 5.4
Swing time (s) 0.41 (0.05) 0.45 (0.05) 0.93 −0.037 (0.020) 0.037 (0.019) 8.3
Stride length (m) 1.43 (0.22) 1.45 (0.22) 0.95 −0.014 (0.067) 0.053 (0.043) 3.6
Velocity (m/s) 1.30 (0.37) 1.31 (0.37) 0.99 −0.012 (0.061) 0.048 (0.040) 3.7