Table 3.

Overview of spatio-temporal gait parameters for eleven healthy subjects and four PD patients (n = 1166 steps). Shown are the mean parameters (SD), Pearson correlation coefficient r, bias (SD), absolute error (SD) and the relative absolute error.

Parameter Sensor Camera r Bias Abs. Error Abs. Error (%)
Stride time (s)
  Healthy 1.13 (0.18) 1.13 (0.18) 1.00 −0.001 (0.015) 0.012 (0.009) 1.1
  Patient 1.27 (0.15) 1.27 (0.15) 0.99 0.003 (0.020) 0.016 (0.013) 1.3
Stance time (s)
  Healthy 0.72 (0.13) 0.69 (0.13) 0.99 0.036 (0.020) 0.037 (0.019) 5.4
  Patient 0.84 (0.12) 0.80 (0.12) 0.99 0.042 (0.020) 0.042 (0.020) 5.4
Swing time (s)
  Healthy 0.41 (0.05) 0.44 (0.05) 0.94 −0.037 (0.019) 0.037 (0.019) 8.2
  Patient 0.43 (0.04) 0.47 (0.04) 0.82 −0.039 (0.026) 0.041 (0.023) 8.5
Stride length (m)
  Healthy 1.45 (0.21) 1.47 (0.21) 0.95 −0.016 (0.066) 0.053 (0.044) 3.6
  Patient 1.25 (0.18) 1.26 (0.17) 0.93 −0.001 (0.065) 0.052 (0.039) 4.2
Velocity (m/s)
  Healthy 1.34 (0.37) 1.35 (0.37) 0.99 −0.013 (0.062) 0.049 (0.041) 3.6
  Patient 1.01 (0.24) 1.02 (0.24) 0.98 −0.004 (0.052) 0.041 (0.031) 4.2