Skip to main content
. 2017 Aug 3;12(8):e0182739. doi: 10.1371/journal.pone.0182739

Table 1. Selection of sample dilution.a.

%CV Low MedLo Med High
ALL 32.12% 24.73% 6.72% 13.14%
1:50, 1:100 5.88% 4.68% 4.76% 4.44%
1:25, 1:50, 1:100 8.05% 6.08% 5.67% 17.41%
1:50, 1:100, 1:200 16.39% 10.50% 5.08% 5.24%
1:100, 1:200, 1:400 26.46% 21.63% 2.28% 4.31%

a Four samples with previously characterized anti-BG-Atri IgM levels of low, medium-low, medium and high values were tested at five different dilutions (1:25, 1:50, 1:100, 1:200 and 1:400). Each result was interpolated to the standard curve, adjusted for dilution, and %CVs were calculated in comparison to all four dilutions or a subset. Two dilutions at 1:50 and 1:100, provided good agreement (after adjusting for dilution factor) across all four samples, with %CV under 6% for all samples.