Table 3.
Test material ID | Number of replicates | NTA 2.3 | NTA 3.0 | ||||
---|---|---|---|---|---|---|---|
Fraction 1 | Fraction 2 | R s | Fraction 1 | Fraction 2 | R s | ||
PSL-50-100_2:1 | 3 | 55 ± 3 | 95 ± 3 | 0.7 | 57 ± 5 | 105 ± 3 | 2.0 |
PSL-50-100_4:1 | 3 | 51 ± 3 | 100 ± 2 | 0.9 | 53 ± 2 | 100 ± 4 | 1.9 |
PSL-60-100_1.2:1 | 4 | 62 ± 5 | 98 ± 3 | 0.7 | 65 ± 3 | 102 ± 4 | 1.4 |
PSL-60-100_2.3:1 | 4 | 63 ± 2 | 94 ± 5 | 0.5 | 65 ± 4 | 97 ± 5 | 1.2 |
PSL-50-80_3.6:1 | 9 | <LOD | 78 ± 2 | N.A. | 53 ± 13 | 83 ± 4 | 1.0 |
PSL-80-100_3.1:1 | 3 | 86 ± 3 | <LOD | N.A. | 84 ± 1 | 100 ± 2 | 0.5 |
PSL-50-60_1.7:1 | 3 | <LOD | 59 ± 2 | N.A. | <LOD | 63 ± 1 | N.A. |
PSL-50-60_5.2:1 | 3 | 48 ± 1 | <LOD | N.A. | 52a | 61a | 0.5a |
N.A. not applicable as only one fraction could be detected
aBased on a single measurement result