Skip to main content
. 2017 Aug 4;19(8):271. doi: 10.1007/s11051-017-3966-8

Table 7.

Relative standard and expanded measurement uncertainties (%) for mean particle size results

Uncertainty component NTA 2.3 NTA 3.0
PSL-46 PSL-100 ERM-FD102 (size class B) PSL-46 PSL-100 ERM-FD102 (size class B)
Precision (u prec) 8.2 4.2 4.2 8.0 5.0 3.5
Trueness (u t) 4.0 4.0 4.0 3.9 3.9 3.9
Expanded (U) 18.2 11.6 11.6 17.8 12.7 10.6