Table 1.
The MM CTC assay is repeatable across titration points. Intra-assay variability was calculated for each titration point from triplicates stained on three separate days. High consistency within the assay (low intra-assay variability, % CV<20) was demonstrated for most conditions. High % CV for n = 1 and 10 was expected due to high variability for detecting low number of cells.
| Run | Expected | Mean | St dev | % CV |
|---|---|---|---|---|
| 1 | 1 | 1 | 0.6 | 43.3 |
| 10 | 9 | 1.5 | 17.6 | |
| 100 | 62 | 9.1 | 14.7 | |
| 1000 | 994 | 26.1 | 2.6 | |
| 2 | 1 | 1 | 1.2 | 173.2 |
| 10 | 5 | 0.6 | 10.8 | |
| 100 | 57 | 5.7 | 10.0 | |
| 1000 | 948 | 46.2 | 4.9 | |
| 3 | 1 | 0 | 0.6 | 173.2 |
| 10 | 7 | 3.5 | 49.5 | |
| 100 | 59 | 8.7 | 14.9 | |
| 1000 | 951 | 23.3 | 2.5 |