Table 1.

Coil maximum 10-g average SAR and power efficiency data

Breast Phantom Tissue Classification Maximum 10-g Average SAR [W/kg/3T2] Input Power [W] +
BI-RADS-a UW Phantom 1 almost entirely fat 0.30 0.63
BI-RADS-a UW Phantom 2 almost entirely fat 0.30 0.64
Uniform fat with skin layer almost entirely fat 0.45 0.67
BI-RADS-b UW Phantom 1 scattered fibroglandular tissue 0.25 0.62
BI-RADS-b UW Phantom 2 scattered fibroglandular tissue 0.30 0.65
BI-RADS-b UW Phantom 3 scattered fibroglandular tissue 0.50 0.69
BI-RADS-c UW Phantom 1 heterogeneous fibroglandular tissue 0.61 0.72
BI-RADS-c UW Phantom 2 heterogeneous fibroglandular tissue 0.58 0.68
BI-RADS-c UW Phantom 3 heterogeneous fibroglandular tissue 0.69 0.72
BI-RADS-d UW Phantom 1 extreme fibroglandular tissue 0.89 0.75
Uniform gland with skin layer extreme fibroglandular tissue 3.4 1.8