Table 3.

Technical validation data of the DCN-CS ELISA assay

Tecnical validation step DCN-CS performance
Detection range (LLOD-ULOD) 1.2–345.3 ng/mL
Lower limit of quantification (LLOQ) 5.3 ng/mL
Intra-assay variation 3%
Inter-assay variation 13%
Dilution of serum samples 1:4
Dilution recovery (1:4 pre-dilution)a 100% (82–113%)
Freeze/thaw recovery (4 cycles)a 94% (90–97%)
Analyte stability up to 24 h, 4 °Ca 87% (86–90%)
Interference Lipids, low/high 107%/86%
Interference Biotin, low/high 100%/100%
Interference Hemoglobin, low/high 100%/100%

aPercentages are reported as mean with range shown in brackets