Skip to main content
. 2017 Jul 1;10(7):733. doi: 10.3390/ma10070733

Table 1.

Clinical information and reasons for revision for all 148 retrieved implants.

Bearing Couple Ceramic on Ceramic Metal on Metal Mann-Whitney U
(p Value)
12/14 Male Taper Material Titanium Alloy Titanium Alloy
Clinical Information (Mean ± SD)
Age at implantation (year) 55.5 ± 12.9 54.2 ± 13.1 0.52
Height (cm) 169 ± 11 167 ± 11 0.32
Weight (kg) 80 ± 19 75 ± 19 0.20
BMI (kg/m2) 27 ± 5 26 ± 5 0.22
Implantation time (year) 5.6 ± 4.6 5.8 ± 3.2 0.32
Reason for Implant Revision (Number of Retrieved Implants)
Aseptic loosening 33 46
Infection 16 11
Periprosthetic femoral fracture 6 5
Adverse local tissue reaction / 21
Pain 2 3
Recurrent dislocation 4 1
Total number of retrieved implants 61 87
Heads including an adapter sleeve / 40 *

* All adapter sleeves were made of cobalt alloy.