Table 1. Comparison of measured volume using three-dimensional ultrasound and estimated volume from 2D diameters (V=43πr1r2r3) with the true volume of tumor phantoms.

Measured Volume
(observer 1)
Measured Volume
(observer 2)
Estimated volume from 2D diameters
(V=43πr1r2r3)
Mean volumea 7.18 ± 5.44 8.31 ± 6.32 9.10 ± 8.47
Mean difference from reference volume (95% CI) -0.24 (-0.79 to 0.32) 0.89 (0.38 to 1.40) 1.69 (0.55 to 2.82)
SD of differences between measured volume and reference volume 1.75 1.58 3.56
Upper limit of agreement 3.19 (2.23 to 4.15) 3.99 (3.12 to 4.86) 8.66 (6.70 to 10.62)
Lower limit of agreement -3.66 (-4.63 to -2.70) -2.21 (-3.08 to 1.34) -5.29 (-7.25 to 3.33)

a The mean reference volume was 7.42 ± 5.76 mL