Table 1.

MR parameters for in vivo study

Scanner 1 Scanner 2


Standard SNR High SNR Standard SNR High SNR
TR [ms] 14.4 12.1 15.1 12
Number of echoes 6 6 6 6
Minimum and maximum TE [ms] 1.2, 11.4 1.1, 10.5 1.2, 11.4 1.1, 10.4
Matrix 256 × 160 128 × 120 224 × 144 128 × 128
Field of view [cm] 42 42 45 45
Slice thickness [mm] 8 10 8 10
Flip angle
Number of slices* 32 24 32 28
Bandwidth 125 50 100 50
Partial ky acquisition 90% 80% 80% 80%
Autocalibrated parallel imaging ×2.65 - ×2.33 -
Voxel size [mm3] 35 115 50 124
Estimated SNR 23.7 ± 2.0 78.3 ± 6.2 52.3 ± 5.1 142.6 ± 9.8

Autocalibrated parallel imaging is expressed as the actual acceleration in acquisition time. signal-to-noise ratio (SNR) was estimated by Monte-Carlo based pseudo-multiple replica method (42). TR, repetition time; TE: echo time.