Skip to main content
. 2017 Aug 7;10(8):609–618. doi: 10.1177/1756283X17722745

Table 3.

Overall concordance in the classification at the 2000 IU/ml and 20,000 IU/ml medical decision points between HPS/CTM HBV v2 and cobas 6800/8800 HBV assays*.

Concordance between HPS and c6800 c6800/8800 HBV Total OPA
(95% CI)
<2000 IU/ml ⩾2000 IU/ml
2000 IU/ml cutoff
HPS/CTM HBV v2 <2000 IU/ml 134 8 142 95.3%
(92.3–97.4%)
⩾2000 IU/ml 6 152 158
Total 140 160 300
Concordance between HPS and c6800/8800 c6800/8800 HBV Total OPA
(95% CI)
<20,000 IU/ml ⩾20,000 IU/ml
20,000 IU/ml cutoff
HPS/CTM HBV v2 <20,000 IU/ml 202 3 205 97%
(94.4–98.6%)
⩾20,000 IU/ml 6 89 95
Total 208 92 300
*

cobas 6800/8800 HBV, cobas HBV test for use on the cobas 6800/8800 Systems; HPS/CTM HBV v2, COBAS TaqMan HBV test for use with the High Pure System.

CI, confidence interval; HBV, hepatitis B virus; OPA, overall percent agreement.