Table 1.

Summary of the FUS+HT PID controller in the 4 treatment groups.

Experimental Group Temperature Response
(Mean ± SD)
Thermal Dose
CEM43°C
Time above 41.3°C
Control (n=3) 36.79 ± 0.15 °C 0.00 ± 0.00 0 ± 0 s
10× 30s FUS+HT to 42°C (n=6) 42.07 ± 0.09 °C 1.04 ± 0.15 221 ± 37 s
3.5min FUS+HT to 42°C (n=4) 41.97 ± 0.06 °C 0.78 ± 0.08 218 ± 4s
20min FUS+HT to 42°C (n=4) 41.99 ± 0.01 °C 4.55 ± 0.20 1287 ± 22s

The temperature response was measured here as the mean temperature reading from the max temperature thermocouple while the prescribed temperature was at the target of 42°C. Thermal dose was calculated as in 70 when both thermocouples read greater than 39°C. The time above 41.3°C was measured as the total time both thermocouples read greater than 41.3°C. Data are presented as the mean ± SEM.