Skip to main content
. 2017 Aug 14;38(8):e224–e231. doi: 10.1097/MAO.0000000000001439

FIG. 3.

FIG. 3

Specifications (A) and photographs of the prototype on table (B) and inserted into the cochlea (C) of the guinea pig cochlear implant with four electrodes (arrows in B). The length of the intracochlear array is 4 mm, as indicated with a stopper (blue oval) located between two dark markers at distance 3 and 5 mm from the tip for better estimate of achieved insertion depth. Catheters for intracochlear injection of the nanomatrix in fluidic form (D). The catheters were manufactured in two variants with different diameters of the silicone tubing (0.3 mm with yellow standard syringe needle and 0.6 mm with white standard syringe needle). The black dot indicates the 3-mm insertion depth as a reference point (arrows in D). CI indicates cochlear implants.