Skip to main content
. 2017 Aug 17;12(8):e0183492. doi: 10.1371/journal.pone.0183492

Table 1. Gelation temperatures of 10%, 15%, and 20% gelatin, measured by rheology and micro-thermomechanical sensors for comparison.

Sensor#1 Sensor#2
Gelation Conc.
(wt%)
Rheology
C)
Average
C)
Error
(%)
Average
C)
Error
(%)
10 28.4 28.33±0.25 0.25 25.57±0.12 2.92
15 30.2 29.37±0.26 2.75 29.33±0.12 2.88
20 31.9 32.40±0.33 1.57 32.87±0.45 3.04