(a) Decomposition of the SMR into equivalent fluidic circuit analog for mesh
current analysis modelling of the dual pressure control integrated with SMR.
Where P represents the inlet pressure of the SMR set by the pressure control
system, i_1 and i_2 represent the fluid flow through the system defined by the
mesh current analysis method, and R represent the defined resistances of the
SMR. (b) Relationship of Reynolds number to the pressure from the mesh current
analysis. (c) Control system performance with Reynolds number within the
synthesis range. (d, e) Performance of our custom feedback pressure control
system (PC) to syringe pump (SP) for (d) transient and (e) steady-state
responses. The size distributions and polydispersity index (PDI) of LPNPs
produced for (f, g) the steady-state period of the following 2.5 minutes
(n=4; *p=0.0221) and (h, i) the transient period of the
beginning 5 seconds (n=3; **p=0.0178). PC is our
pressure control system and SP is a commercial syringe pump. Significance was
determined by a two-tailed unpaired t-test.