. 2017 Aug 1;30(4):223–229. doi: 10.1089/jamp.2016.1333

Table 2.

Device Output and Delivery Comparison

Device Emitted rate (μL/min) Expected lung deposition rate (μL/min)a NaCl deposition rate (mg/min)b NaCl deposited per day (mg)
tPAD 24 (9) 9.4 0.67 384c
PARI LC Star 121 (13) 43.6 3.1 151d

Mean values (SD) for output and anticipated lung delivery with tPAD and LC Star devices.

a

Calculation utilizes the deposition fraction for each device that was measured in this study (39% of emitted dose from tPAD and 36% of emitted dose from LC Star).

b

Calculation made for 7% NaCl solution.

c

Presumes tPAD is used for 8 hours (e.g., overnight).

d

For PARI LC Star, presume that 3 mL of the 4 mL loaded dose is nebulized (i.e., 1.0 mL dead volume), that the device is used twice daily, and that 36% of this emitted dose deposits in the lung.