Published in final edited form as: J Prosthet Dent. 2014 May 14;112(4):817–823. doi: 10.1016/j.prosdent.2014.03.016

Table I.

Comparison of total number of misfits, no misfit, and unsure answers obtained from radiographs made with or without PIXRL device

Misfit Size Misfit (%) No Misfit (%) Unsure (%) Total Pa
0 μm .364
 No PIXRL 7 (8.6) 57 (70.4) 17 (21.0) 81
 PIXRL 4 (4.9) 64 (79.0) 13 (16.1) 81
 Total 11 121 30 162

50 μm <.001
 No PIXRL 13 (16.1) 50 (61.7) 18 (22.2) 81
 PIXRL 63 (77.8) 13 (16.0) 5 (6.2) 81
 Total 76 63 23 162

100 μm .049
 No PIXRL 75 (92.6) 2 (2.5) 4 (4.9) 81
 PIXRL 81 (100) 0 (0) 0 (0) 81
 Total 156 2 4 162

PIXRL, Precision Implant X-ray Relator and Locator.

a

Cochran-Mantel-Haenszel test was used to compare ratings between radiographs with or without PIXRL device, conditional on raters (clinicians).