Table 1.

DRS model discrepancies between the reference data and calculation for the default beam model and the final customized model for the Varian base class. Reference dosimetric parameters were determined previously.8

PDD Jaw Output Factor

Default Tuned Default Tuned
5 cm −0.1% −0.1% 6×62 cm 0.9% 0.2%
10 cm −0.2% −0.2% 15×15 cm2 −0.3% 0.0%
15 cm 0.6% 0.2% 20×20 cm2 −0.2% 0.0%
20 cm 0.3% −0.8% 30×30 cm2 0.3% −0.1%
IMRT - style Output Factors SBRT - style Output Factors

Default Tuned Default Tuned
6×6 cm2 0.4% 0.1% 6×6 cm2 1.0% 0.0%
4×4 cm2 −0.3% −0.8% 4×4 cm2 1.3% −0.1%
3×3 cm2 −0.2% −0.8% 3×3 cm2 1.7% 0.0%
2×2 cm2 −0.7% −1.2% 2×2 cm2 2.1% −0.4%
Off - Axis Factors
Default Tuned
5 cm −0.6% −0.1%
10 cm −0.2% 0.0%
15 cm −0.4% 0.0%