Skip to main content
. Author manuscript; available in PMC: 2017 Aug 23.
Published in final edited form as: Anal Bioanal Chem. 2016 Oct 28;409(4):931–937. doi: 10.1007/s00216-016-9933-x

Table 4.

Precision, limits of detection (LODs), and upper linearity limits.

Analytes QCH* QCL* LOD (ng/mL) Upper linearity limit (ng/mL)
Mean (ng/mL) Within-run precision (CV%) Method precision (CV%) Mean (ng/mL) Within-run precision (CV%) Method precision (CV%)
1-OH-NAP 7.78 3.3% 4.8% 1.22 5.9% 7.4% 0.06 200
2-OH-NAP 8.94 3.2% 5.7% 1.54 4.4% 6.2% 0.09 200
2-OH-FLU 1.06 3.8% 5.4% 0.25 4.3% 5.7% 0.008 25
3-OH-FLU 1.14 3.3% 4.8% 0.28 4.2% 5.1% 0.008 10
1-OH-PHE 0.68 7.2% 8.0% 0.17 8.6% 9.7% 0.009 10
Σ2,3-OH-PHE 1.36 6.5% 7.9% 0.36 7.1% 8.2% 0.01 20
4-OH-PHE 0.54 7.0% 10.5% 0.13 9.5% 12.6% 0.007 10
1-OH-PYR 0.79 12.1% 12.4% 0.40 11.4% 13.0% 0.07 10
*

QCH: QC high; QCL: QC low.