TABLE II.

Light Intensity Setting on the LCU and Corresponding Intensity Reading on the Visible Light Intensity Meter

Light Intensity on the LCU (mW/cm2) Light Intensity Read by the Visible Light Intensity Meter (mW/cm2)
300 229
550 455
800 679