Repeatability testing of PT/INR measurement with LSR sensor.

Normal PT/INR High PT/INR
PTLAB 13.4 27.2
INRLAB 1.08 2.52
PTLSR, mean ± SD (%CV) 13.7 ± 0.6 (4.4%) 28.0 ± 1.4 (4.9%)
INRLSR, mean ± SD (%CV) 1.08 ± 0.05 (5.2%) 2.52 ± 0.14 (5.7%)

(n = 10 measurements per blood sample, %CV was calculated by dividing the SD by mean).