Table 6. Intra-day (n = 8) accuracy and precision for TH, TAM and TAc in spiked lysed PCCL3 cell extracts in 6-well plates.
0.10 | 12.5 | 40.0 | |||||||
---|---|---|---|---|---|---|---|---|---|
Compound | Mean Calculated Concentration | % Relative Standard Deviation | % Accuracy | Mean Calculated Concentration | % Relative Standard Deviation | % Accuracy | Mean Calculated Concentration | % Relative Standard Deviation | % Accuracy |
T0 | 0.10 | 8.6 | 103 | 12.7 | 3.8 | 102 | 39.8 | 1.4 | 99 |
3T1 | 0.10 | 12.6 | 99 | 12.5 | 3.2 | 100 | 39.3 | 2.2 | 98 |
3'T1 | 0.09 | 12.7 | 94 | 12.8 | 2.8 | 103 | 40.1 | 1.1 | 100 |
3,5T2 | 0.10 | 5.3 | 100 | 12.9 | 2.4 | 103 | 39.3 | 2.2 | 98 |
3,3'T2 | 0.10 | 9.1 | 98 | 12.8 | 1.2 | 102 | 40.5 | 1.3 | 101 |
3',5'-T2 | 0.11 | 6.6 | 108 | 12.6 | 2.4 | 101 | 39.9 | 1.9 | 100 |
T3 | 0.11 | 8.6 | 106 | 12.7 | 1.9 | 102 | 40.4 | 1.8 | 101 |
rT3 | 0.11 | 7.0 | 106 | 12.2 | 1.5 | 97 | 38.7 | 3.6 | 97 |
T4 | 0.10 | 4.5 | 103 | 12.6 | 1.6 | 101 | 39.9 | 1.6 | 100 |
T0AM | 0.09 | 14.0 | 89 | 12.5 | 1.2 | 100 | 39.0 | 1.7 | 98 |
3T1AM | 0.10 | 5.6 | 95 | 12.4 | 0.8 | 99 | 39.9 | 1.0 | 100 |
3,5T2AM | 0.09 | 7.9 | 94 | 12.1 | 3.2 | 97 | 39.5 | 1.5 | 99 |
T3AM | 0.10 | 8.8 | 104 | 11.1 | 17.0 | 89 | 39.4 | 3.2 | 98 |
rT3AM | 0.10 | 6.3 | 101 | 11.1 | 18.8 | 89 | 38.4 | 4.2 | 96 |
T4AM | <LLOQ | 11.4 | 19.4 | 91 | 36.9 | 6.1 | 92 | ||
T0Ac | <LLOQ | 12.6 | 2.5 | 101 | 39.3 | 3.5 | 98 | ||
T1Ac | <LLOQ | 13.6 | 4.4 | 109 | 40.2 | 3.7 | 101 |