Ionic current, conductance, and rectification of truncated AQP system.
(a–c) Ionic current across truncated AQP embedded in a lipid membrane,
Figure 1c,d, at
±500 mV, ±800 mV, and
±1200 mV biases respectively, with magnitude of
bias given as —V— at the top of each panel. Data at
±1000 mV are reported in Figure S1. Purple traces indicate
current whereas step-function black traces indicate the applied transmembrane
bias versus simulation time. The ionic current traces show 2 ns
running average of the instantaneous current sampled every 9.6 ps. Dashed lines
serve as guides for the eye. Positive current travels from cis
to trans as defined in Figure
1c. (d,e) Mean ionic current (d) and ion conductance (e) of the
truncated AQP tetramer as a function of transmembrane bias. Mean current is
calculated by, first, finding the time-average of the ionic current across the
tetramer for each constant bias fragment of the trajectory. The weighted average
of these currents is the mean current reported. Error bars in both panels
indicate the standard deviation of the mean. The mean conductance and its
standard deviation are found by dividing mean current by magnitude of applied
bias. The inset of (e) shows the ratios of the current at negative biases to the
current at the corresponding positive biases. Error bars specify the propagated
standard deviations in the mean currents.