Skip to main content
. 2017 Feb 13;44(3):847–860. doi: 10.1002/mp.12083

Table 2.

Estimated escape efficiency and light output (keV−1) for interactions at the back (z = 0) and front (z = L) of the investigated samples. Light output values are determined by dividing escape efficiency estimates by W = 17 eV

Back Front
ε¯(0)
keV−1
ε¯(L)
keV−1
FOS 150 HR 0.073 4.3 0.228 13.4
FOS 600 HR 0.069 4.1 0.207 12.2
FOS 200 HL 0.222 13.1 0.313 18.4
FOS 600 HL 0.201 11.8 0.311 18.3
FOS 1000 HL 0.176 10.4 0.293 17.2
GPXS 200 0.600 35.3 0.810 47.6
GPXS 600 0.575 33.8 0.814 47.9
GPXS 1000 0.501 29.5 0.884 52.0