Skip to main content
. 2017 Mar 8;35(8):1831–1837. doi: 10.1002/jor.23464

Table 1.

Deflection Angles of Spinal Implants in 3 T and 7 T Fields

Spinal Implant Averaged Deflection Angle (°)
Material Length (cm) Weight (g) 7 T 3 T
Rod
Pure titanium 5 5.5 5.0 2.0
10 10.8 5.8 1.7
15 16.2 6.2 1.0
20 21.6 6.2 1.0
20,  cross‐linked 66.0 N.A. N.A.
Titanium alloy 5 5.3 5.7 2.3
10 10.6 6.5 2.3
15 15.9 6.8 1.7
20 21.1 7.7 1.0
20, cross‐linked 65.0 N.A. N.A.
Cobalt–chrome 5 9.9 17.8 6.0
10 19.9 19.2 6.3
15 29.8 20.3 5.7
20 39.9 21.0 5.0
20, cross‐linked 102.6 N.A. N.A.
Range (median) 5.0–20.0 (12.5) 5.5–102.6 (19.9) 5.0–21.0 (6.65) 1.0–6.3 (2.15)
Screw
Titanium alloy/cobalt–chrome 5.8 12.1 10.0 3.2
Cross‐linking bridge
Titanium alloy 5.2 11.4 6.7 2.2