. 2017 Mar 8;35(8):1831–1837. doi: 10.1002/jor.23464

Table 3.

Temperature Changes During MRI Sequence(3D‐GRE) Performed Using 7 T Field

Spinal Implant RF‐Radiation‐Induced Heating (7 T) 3D‐bGRE (15 min)
Material Length (cm) Weight (g) Implant (Δ°C) CTRL‐l (Δ°C) Diff‐1 (°C) CTRL‐2 (Δ°C) Diff‐2 (°C)
Rod
Pure titanium 5 5.5 0.7 0.4 0.3 0.5 0.2
10 10.8 0.7 0.8 −0.1 0.5 0.2
15 16.2 0.6 0.8 −0.2 0.9 −0.3
20 21.6 0.7 0.7 0 0.8 −0.1
20, cross‐linked 66.0 0.6 0.6 0 0.3 0.3
Titanium alloy 5 5.3 1.0 0.6 0.4 0.5 0.5
10 10.6 0.7 0.7 0 0.8 −0.1
15 15.9 0.7 0.7 0 0.6 0.1
20 21.1 0.6 0.8 −0.2 0.7 −0.1
20, cross‐linked 65.0 0.7 0.6 0.1 0.8 −0.1
Cobalt–chrome 5 9.9 1.0 0.9 0.1 0.5 0.5
10 19.9 0.7 0.4 0.3 0.8 −0.1
15 29.8 0.8 0.4 0.4 0.9 −0.1
20 39.9 0.8 0.5 0.3 0.5 0.3
20, cross‐linked 102.6 0.6 0.8 −0.2 0.4 0.2
Range (median) 5–20 (12.5) 5.5–102.6 (19.9) 0.6–1.0 (0.7) 0.4–0.9 (0.7) −0.2 to 0.4 (0) 0.3–0.9 (0.6) −0.3 to 0.5 (0.1)
Screw
Titanium alloy/cobalt–chrome 5.8 12.1 0.7 0.8 −0.1 0.8 −0.1
Cross‐linking bridge
Titanium alloy 5.2 11.4 0.6 0.7 −0.1 0.7 −0.1

CTRL‐1, change in temperature 5 cm away from implant; CTRL‐2, change in temperature after implant removal; Diff‐1, difference of implant from CTRL‐1; Diff‐2, difference of implant from CTRL‐2; Implant, change in temperature of implant.