Skip to main content
. 2015 Dec 24;4:41–49. doi: 10.1016/j.plabm.2015.12.005

Table 2.

Performance of the StatStrip-Lactate and iSTAT-CG4+ tests to predict acidosis in umbilical cord blood at different Lactate cut-offs and pH levels.

Lactate>6.6 mmol/L
Lactate>5.1 mmol/L
pH<7.20 pH<7.05 pH<7.20 pH<7.05
StatStrip-Lactate
 Sensitivity (%) 25 60 46 80
 Specificity (%) 83 97 83 79
 PPVa (%) 92 92 96 70
 NPVb (%) 12 80 16 87
iSTAT-CG4+
 Sensitivity (%) 35 70 52 80
 Specificity (%) 83 88 83 71
 PPVa (%) 94 78 96 62
 NPVb (%) 14 83 18 86
a

PPV, positive predictive value.

b

NPV, negative predictive value.