Skip to main content
. 2015 Nov 27;4:1–10. doi: 10.1016/j.plabm.2015.11.001

Table 5.

Results obtained with the TSH Beckman Coulter Access® assay, both in the normal format (with the reagent kit provided by the manufacturer) and with the ‘nonsense’ in-house format (see text). Patient case is the patient described in the text. Other patients studied are identified by laboratory ID number.

Sample Normal TSH assay (µU/mL) ‘Nonsense’ TSH assay (µU/mL)
QC low level 0.65 <0.003
QC high level 25.6 0.12
Patient case 7.49 8.38
Patient 1134320 3.60 0.01
Patient 6461018 1.12 <0.003