Table 5.
Results obtained with the TSH Beckman Coulter Access® assay, both in the normal format (with the reagent kit provided by the manufacturer) and with the ‘nonsense’ in-house format (see text). Patient case is the patient described in the text. Other patients studied are identified by laboratory ID number.
Sample | Normal TSH assay (µU/mL) | ‘Nonsense’ TSH assay (µU/mL) |
---|---|---|
QC low level | 0.65 | <0.003 |
QC high level | 25.6 | 0.12 |
Patient case | 7.49 | 8.38 |
Patient 1134320 | 3.60 | 0.01 |
Patient 6461018 | 1.12 | <0.003 |