Skip to main content
. 2017 Jan 4;7:19–26. doi: 10.1016/j.plabm.2016.12.002

Table 1.

Within-run and total imprecision using the CLSI standard EP5A2.

Siemens EMIT® MPA controls
Within-run
Total
IQC levels Days Mean (mg/L) CV (%) 95% CI CV (%) 95% CI
1 20 1.2 5.5 4.5–6.9 9.3 7.4–11.3
2 20 9.3 1.2 1.0–1.6 4.1 2.9–5.3
3 20 14.1 1.5 1.2–1.9 2.5 1.8–3.2
Roche Total MPA controls
Within-run
Total
IQC levels Days Mean (mg/L) CV (%) 95% CI CV (%) 95% CI
1 20 0.8 4.3 3.5–5.5 9.7 8.0–12.1
2 20 3.5 1.3 1.1–1.6 2.5 2.1–3.1
3 20 12.6 1.9 1.6–2.5 2.9 2.4–3.5