Table 1.

Manufacturer specifications (package insert) for the direct bilirubin methods.

Method Hemolysis index limit Lipemia index limit Linearity of method (mg/dL) Dilution protocol Reference range (mg/dL)
Roche diagnostics 30 60 0.2–10.0 Manual 0.0–0.3
Diazyme laboratories 500 Not listed 0.1–20.0 Auto-dilution 0.0–0.2
Randox laboratories 1000 1000 0.1–15.4 Auto-dilution 0.0–0.2